Digital Ohmmeter
If you are interested in this product,
you can
Product Name:
Digital Ohmmeter
Product Model:
DO6
Supplier Name:
shanghai beihan
Brief Introduction
• DO6 Digital Ohmmeter
• Resolution of 100 nano ohm on 2 milli ohm range
• Manual / automatic range selection
• Temperature compensation for copper and brass
Digital Ohmmeter introduction
Digital Ohmmeter
Model:DO6
FEATURES
Covering a wide range of measurements from 1 micro ohm to 200 kilo ohm, the DO6 ohmmeter is a comprehensive ohmmeter with many features incorporated as standard. These benefits include manual and automatic range selection; temperature compensation for copper, referencing the measurement of copper to 20°C; voltage limiting facility for those wishing to measure switch contact resistance to International Standards, the measurement voltage is limited to 20mV. An RS232 serial interface gives full talk / listen capability, making it simple to add to an automated measuring system. Hi / Lo limits may be set over the interface with voltage-free contacts being available on the rear panel. For those wishing to measure inductive circuits e.g. motor windings, a special protection circuit may be initiated thus avoiding any problems that may arise due to induced high voltage.
Key Features
• True 4 wire measurement eliminates lead resistance errors
• 2 push button ranges 2 milli ohm… 200 kilo ohm
• Resolution of 100 nano ohm on 2 milli ohm range
• Manual / automatic range selection
• Temperature compensation for copper and brass
• Thermal E.M.F. compensation
• Dry circuit measurement: 18mV limited
• Open circuit lead warning
• RS232 interface
• Hi / Lo limits
Technical Specification
|
Measurement
|
4 Terminal Kelvin/Thomson, eliminating lead resistance errors
|
|
Display
|
4.5 digit 15mm 7 segment LED with polarity and overload indication. 2 digit 16 segment LED for unit display
|
|
Ranges
|
9 ranges manual selection with up/down keys, automatic selection with autorange function
|
|
Auto Zero
|
Permits the automatic zero of amplifiers and external circuits
|
|
Temperature Compensation
|
Measurement compensated for the following referenced to 20°C:Copper Cu = 3.93%/KBrass Ms63 = 1.50%/KBrass Ms80 = 1.60%/KAmbient sensed with Pt 100 sensor
|
|
Relay Output
|
For the pre-selected upper and lower limit values “Too small”,“Correct”, “Too large”Contact load = 30W max 48V 1 Amp
|
|
Dry Circuit Measurement
|
In accordance with DIN 41640 part 4 and IEC – 132-1Open circuit voltage limited to 18mV
|
|
Interface
|
RS232 25 pin D-Bush talk listen function
|
|
Terminals
|
4mm safety sockets
|
|
Working Temperature
|
0°C to 50°C rel. humidity 90% max. non-condensing
|
|
Storage Temperature
|
-20°C to +60°C
|
|
Mains Supply
|
230 Volts +6% - 10%45Hz to 65Hz approx. 30VA
|
|
Safety
|
IEC 1010 Protective Class 1
|
|
Dimensions
|
255mm x 90mm x 263mm (W D H) approx
|
|
Mass
|
5kg approximately
|
Resistance Ranges
|
Range
|
Resolution
|
Typical Current
|
Accuracy@20°C±5°C/ 1 year
|
Temperature Coefficient/°C
|
|
20.000mΩ
|
1µΩ
|
1A
|
±0.05% FS
|
<50ppm
|
|
200.00mΩ
|
10µΩ
|
100mA
|
±0.05% FS
|
<50ppm
|
|
2000.0mΩ
|
100µΩ
|
10mA
|
±0.05% FS
|
<50ppm
|
|
20.000Ω
|
1mΩ
|
10mA
|
±0.05% FS
|
<50ppm
|
|
200.00Ω
|
10mΩ
|
1mA
|
±0.05% FS
|
<50ppm
|
|
2.0000kΩ
|
100mΩ
|
1mA
|
±0.05% FS
|
<50ppm
|
|
2.0000kΩ
|
1Ω
|
100µA
|
±0.05% FS
|
<50ppm
|
|
2.0000kΩ
|
10Ω
|
10µA
|
±0.05% FS
|
<50ppm
|
Leave a message for product